GET THE APP

Assessment of the analytical quality of the electrolyte anal | 18250
International Research Journals

International Research Journal of Biochemistry and Bioinformatics

All submissions of the EM system will be redirected to Online Manuscript Submission System. Authors are requested to submit articles directly to Online Manuscript Submission System of respective journal.

Assessment of the analytical quality of the electrolyte analyzer i-Smart 30 PRO for the measurement of serum electrolytes

Abstract

Sekossounon Sanni

The i-Smart 30 PRO electrolyte analyzer is an electrochemical analyzer used to measure the concentrations of sodium, potassium and chlorides in various biological fluids. The aim of this study was to assess the analytical quality of the electrolyte analyzer i-Smart 30 PRO and validate its use for electrolytes testings at the Laboratory of Clinical Biochemistry and Toxicology, National Teaching Hospital Hubert Koutoukou MAGA. Samples received routinely for sodium, potassium and chlorides testings were runned on the electrolyte analyzers i-Smart 30 PRO and ILyte used as the gold standard. The sodium, potassium and chlorides serum concentrations were biased with significant differences. However more than 95% of the testings results were within the 95% limits of agreement. The i-Smart 30 PRO electrolyte analyzer could validly replace the ILyte analyzer for the measurements of these electrolytes concentrations in serum.

Share this article